A Low-Cost RF Detector to Enhance the Direct Power Injection Conducted Immunity Setup - LAAS - Laboratoire d'Analyse et d'Architecture des Systèmes Accéder directement au contenu
Article Dans Une Revue IEEE Letters on Electromagnetic Compatibility Practice and Applications Année : 2022

A Low-Cost RF Detector to Enhance the Direct Power Injection Conducted Immunity Setup

Alexandre Boyer

Résumé

Direct power injection (DPI) is a convenient method to characterize the conducted susceptibility of integrated circuits (IC). However, a practical issue of this test is that the voltage, current and impedance of the tested pin remain unknown during the test, even though it can provide valuable information to IC designers about IC failures. This letter presents an enhancement of the standard DPI test based on a commercial and affordable RF detector, which covers the frequency range of 5 MHz to 3 GHz. The proposed approach does not require specific measurement probes and board design constraints. It can be easily inserted in a conventional DPI test bench without any influence on the test results.
Fichier principal
Vignette du fichier
Boyer 2022 - Enhancement DPI setup IC immunity_final.pdf (1.48 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03794115 , version 1 (03-10-2022)

Identifiants

Citer

Alexandre Boyer. A Low-Cost RF Detector to Enhance the Direct Power Injection Conducted Immunity Setup. IEEE Letters on Electromagnetic Compatibility Practice and Applications, 2022, 4 (4), pp.108 - 113. ⟨10.1109/LEMCPA.2022.3210876⟩. ⟨hal-03794115⟩
21 Consultations
82 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More